Zum Hauptinhalt springen
Zurück zu Datenmodellen

INSTR-CORE SCHEMA

Keithley 2602B

This page exposes the command, limit, state, and safety contract that instr-core can load for validation and dry-run planning.

REGISTRY KEY

instr-core lookup contract

instrument_key
keithley/smu/2602b
public entity ID
instr:keithley/smu/2602b
schema version
1.0.0
schema status
derived
record version
0.1.0-demo
record status
unverified

VALIDATION CONTRACT

Limits and commands

These fields are shaped to match instr-core Python InstrumentSchema model.

Global limits

voltage
max 40.4 V
current
max 3.03 A
power
max 40.4 W

Commands

SCPIRangeStateSafety
:SOUR:FUNC

Set the source function (VOLT or CURR)

nonesource_modenone
:SOUR:VOLT

Set the source voltage level

-40.4...40.4:SOUR:VOLTnone
:SOUR:CURR

Set the source current level

-3.03...3.03:SOUR:CURRnone
:SOUR:VOLT:RANG

Set the source voltage range

0...40.4:SOUR:VOLT:RANGnone
:SOUR:CURR:RANG

Set the source current range

0...3.03:SOUR:CURR:RANGnone
:SENS:CURR:PROT

Set the current compliance (protection) level

0...3.03:SENS:CURR:PROTnone
:SENS:VOLT:PROT

Set the voltage compliance (protection) level

0...40.4:SENS:VOLT:PROTnone
:OUTP

Turn the output ON or OFF

noneoutput2 rule(s)
:OUTP:DEL

Set output delay time

nonenonenone
:READ?

Return the measurement from the active source function

nonenonenone